What does Light Transmission (LT) measure?

LightControl parameter LT.

  • What does it measure?
    • Light troughput of the lenses.
  • How does it measure it?
    • RGB value converted to luminance.
  • How does it relate to the previous measurement?
    • It uses an average reference value from a group of new scopes.

Broader explanation

This measurement is done to ensure good visibility and light throughput for the surgeon during operation.

The measuring device uses a calibrated light source which passes through the endoscope to the eyepiece which is connected to a camera.

The camera takes consecutive images and the system records relevant data on a pixel level.

Each pixel has an RGB value in the color spectrum. The combination of all pixels is converted into a luminance value by an algorythm specifically designed for this task.

The system uses the images to cross-reference the measured value with data known about this brand and type of endoscope in the system using a bespoke algorythm.

Based on the measured value and the reference value, a score between 0 and 100% is given. The pass/fail score can be adjusted by the user. Dovideq sets it at 70% as a standard.