How long and at what pressure does LeakControl test?

Set your own conditions for leak control.

Test time:
When you first receive your LeakControl it will be set to 60 seconds, but this can be changed by the user if so desired. The measuring time can be set to 60 seconds, 120 seconds or auto detect.

When set to auto detect, the LeakControl will determine the scope volume and test for 60 seconds for small scopes, and 120 seconds for larger scopes.

 

System pressure:
The LeakControl pressurizes the scope to a pressure of 220mbar.
When there is a leak in the scope, the LeakControl will keep the scope pressurized to prevent ingress of liquids.

In the unfortunate event of a LeakControl failure, the pressure will never exceed 370 mbar. This is due to the built-in safety valve that releases pressure above this point. This is to prevent damage to the scope.