How to set the LeakControl pressure transmitter output

Get a maximum resolution on the pressure transmitter

Step 1:
Go to the settings menu on your LeakControl.

Step 2:
Go to "Sensor calib". 

Step 3:
System is asking for atmospheric pressure. Make sure the inlet/outlet is clear. Pressure transmitter is at its lowest setting. (e.g. 264)
Press START to confirm.

Step 4:
Connect the calibration set to the LeakControl.
Set the pressure at 300mmhg. Adjust pressure transmitter first to max.(e.g. 844) and then screw one resolution step back. (e.g. 840)
Press START to confirm.
The pressure range is now divided over the maximum possible clipping range of +/- 550 steps.

Step 5:
Repeat the steps one more time to check if the values ​​have been saved.